Component | Description |
---|---|
PyroMark Q24 | Sequencing instrument for quantitative sequence analysis |
PyroMark Q24 Advanced Software | Analysis software with 4 analysis modes (AQ: allele quantification; SNP: genotyping; CpG: methylation at CpG or CpN sites; and SEQ: base-calling of unknown sequences) |
PyroMark Q24 Advanced Reagents | Enzymes, substrates, nucleotides, and buffers. Compatible only with assays set up and analyzed with PyroMark Q24 Advanced Software. |
PyroMark Q24 Advanced CpG Reagents | Enzymes, substrates, nucleotides, and buffers for assays requiring longer sequence read lengths. Compatible only with assays set up and analyzed with PyroMark Q24 Advanced Software. |
PyroMark Q24 Vacuum Workstation | Workstation for preparing up to 24 samples in parallel from PCR product to single-stranded template |
PyroMark Q24 Control Oligo | Control for verification of proper installation and operation of the system |
PyroMark Q24 Advanced | PyroMark Q24 | |
---|---|---|
Throughput | 1–24 samples | 1–24 samples |
Running volume | 25 µl | 25 µl |
PCR requirements | 5–10 µl (~0.5–3 pmol of product) | 5–10 µl (~0.5–3 pmol of product) |
Read length (assay- and sequence-dependent) | 10–140 bp or more | 10–80 bp |
Application software | PyroMark Q24 Advanced SW (requires firmware 1.5.6903 or higher) | PyroMark Q24 SW 2.0 |
Software functionality | SEQ (de novo sequencing) CpG/CpN methylation SNP AQ | SQA (de novo sequencing) CpG methylation AQ/SNP |
Main applications | Complex mutation analysis Epigenetics (CpG and CpN analysis) Resistance typing and microbial ID | Mutation analysis Resistance typing |
Compatible reagents | PyroMark Q24 Advanced Reagents PyroMark Q24 Advanced CpG Reagents | PyroMark Q24 Gold Reagents |
Sensitivity | 2% mutation 98% wt | 2% mutation 98% wt |
Features | Specifications |
---|---|
Applications | Methylation analysis (CpG and CpN sites), allele quantification, genotyping, sequence analysis, microbial typing |
Power | 100–240 V AC, 47–63 Hz, 1.1–0.45 A (grounded); from external power supply to the instrument : 12 VDC and 24 VDC nominal |
Chemical resistance | pH 4 to pH 9, common detergents, 0.5 M sodium hydroxide, ethanol |
Kits designed for this instrument | PyroMark Q24 Advanced Reagents, PyroMark Q24 Advanced CpG Reagents |
Connections | One USB port (2.0) |
Instrument dimensions | 420 x 390 x 525 mm (16.5 x 15.4 x 20.7 in) |
Place of operation | For indoor use only |
Operating temperature | 15–32°C (59–90°F) |
Pollution level | 2 |
Humidity | Relative humidity of 20–90% (noncondensing) |
Process temperature | 28°C (82.4°F) ± 1% |
Overvoltage category | II |
Process time | Depends on the number of dispensations |
Samples per run (throughput) | 1–24 |
Software | PyroMark Q24 Advanced Software (to be installed on PC) |
Technology | Pyrosequencing |
Weight | 27.5 kg (60.6 lb) |
Altitude | Up to 2000 m (6500 ft) |