Looking for a quick way to design experiments?
Try the Workflow Configurator. A convenient tool to build experimental workflows and find products to match your needs.

careHPV Test Kit

For testing for high-risk human papillomavirus (HPV) in limited-resource settings

Products

The use of the careHPV Test is indicated as a primary screening test in women 30 years and older to detect high-risk HPV infection. CARE is a registered trademark of COOPERATIVE FOR ASSISTANCE AND RELIEF EVERYWHERE, INC. (“CARE”). CARE and the members and affiliates of CARE International are not affiliated with QIAGEN and do not sponsor, endorse, support, participate in or control the development, manufacture, use or sale of any QIAGEN product.
Image

careHPV Test Kit (96)

Cat. No. / ID:   614015

For 96 reactions: Assay Microplate, Negative Calibrator, Positive Calibrator, Reagent 1, Indicator Dye, Stabilized biologics (Reagent 2, Reagent 3, Reagent 4, Reagent 6), Reconstitution diluents (Reagent 2 Diluent, Reagent 3 Diluent, Reagent 4 Diluent, Reagent 6 Diluent), Reagent 5, Reagent 5 Nozzle
MX$11,285.00
Image

careHPV Training Panel

Cat. No. / ID:   619029

8 samples of known HPV DNA concentrations and 8 careHPVTraining Brushes

Features

  • Detection of 14 high-risk HPV types
  • Convenient test results in 2.5 hours for possible same-day follow up
  • Accurate, robust and accessible primary screening
  • High clinical sensitivity and specificity

Product Details

The careHPV Test is a signal-amplification, rapid batch diagnostic test for high-risk HPV DNA detection. The careHPV Test allows the qualitative detection of 14 high-risk types of HPV DNA in cervical and vaginal specimens in limited-resource settings.

Performance

A multi-center clinical study using the careHPV Test with the careHPV Test System was conducted with cervical specimens collected using the careBrush and careHPV Collection Medium from women (30–59 years) in a general screening population and outpatients clinics. A total of 1279 women were enrolled in this study, representing a relatively equal distribution across 3 hospitals; 1241 participants completed the study. The 3 study sites collected specimens from approximately 147 patients diagnosed with cervical cancer or precancerous lesions (CIN 2/3+), 162 patients with benign lesions (inflammation/mild cervical intraepithelial neoplasia, CIN 1) and 932 cases of normal control.

Acetic acid staining was also performed for visual examination (VIA). Liquid-based cytology was performed at each hospital and the results recorded using the Bethesda Classification. The careHPV Test, the digene HC2 High-Risk HPV DNA Test and amplification of HPV nucleic acid using a PCR-based fluorescent detection kit were performed for each patient specimen. All careHPV testing was performed at room temperature (15–30°C). Test results were compared to the disease status of each patient. The concordance between the careHPV Test and digene HC2 High-Risk HPV DNA Test was 93.71%. The concordance between the careHPV Test versus PCR-based HPV detection was 90.89%.

In studies comparing self-collected versus physician-collected specimens, the results of careHPV testing indicated reduced but similar sensitivity for CIN 2+: 70–83% for self-collected versus 82–96% for physician-collected. Specificity results were similar for CIN 2+ for both methods: 87–91% for self-collected versus 83–92% for physician-collected specimens.

Studies indicate that the careHPV Test does not cross-react with microorganisms likely to be found in human cervical or vaginal specimens. The careHPV Test does not cross-react with human genomic DNA. There are no overlapping, cross-reactive DNA sequences in HIV, HBV, EBV, CMV, Adenovirus 2 and Neisseria meningitides.
Substances that may be found in cervical or vaginal specimens (whole blood, douche, anti-fungal cream, contraceptive jelly and vaginal lubricant) were tested for impact on the performance of the careHPV Test. False-positive results were observed with anti-fungal cream, but no false positive results were observed with any of the other substances at any concentration tested.

Principle

The careHPV Test utilizes the same Hybrid Capture 2 (HC2) technology developed for QIAGEN digene HC2 High-Risk HPV DNA Test (HC2 Test).

The careHPV Test detects the presence of the high-risk carcinogenic HPV types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68 using full genome probes complementary to HPV DNA, specific antibodies, signal amplification and chemiluminescent detection. The careHPV Test is used with the careHPV Test system to analyze HPV DNA high-risk groups in cervical and vaginal specimens collected with a careBrush and careHPV Collection Medium.

Procedure

The careHPV Test and careHPV Test System detect the presence of 14 high-risk carcinogenic HPV types using full genome RNA probes complementary to the HPV DNA, specific antibodies and chemiluminescent detection. The target DNA combines with specific RNA probes, creating RNA:DNA hybrids. Then, the RNA:DNA hybrids are captured onto a solid phase coated with universal capture antibodies specific for RNA:DNA hybrids. The specimen matrix is washed from the captured hybrids to remove inhibitors. During the signal amplification, captured RNA:DNA hybrids are detected with multiple antibodies conjugated to alkaline phosphatase. The signal resulting from the chemiluminescent reaction is read by the careHPV Test Luminometer. The results are automatically interpreted by the careHPV Test System and are displayed graphically on the careHPV Test Controller screen.

Applications

The use of the careHPV Test is indicated as a primary screening test in women 30 years and older to detect high-risk HPV infection. The HPV types detected by the test are the high-risk HPV types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68. High-risk HPV infection is a risk factor for developing high-grade cervical intraepithelial neoplasia (CIN 2/3+). Cervical specimens for testing may be collected by physicians or trained medical professionals. Vaginal specimens may be self-collected under the guidance of a medical professional. The careHPV test is a cost-effective primary screening tool for cervical cancer in developing countries.

The careHPV Training Panel is used to demonstrate the operation and maintenance of the careHPV Test System, conduct a detailed overview of the careHPV Test procedure, demonstrate each step of the careHPV Test and perform validation of trainee test performance.

Resources

Kit Handbooks (1)
Safety Data Sheets (1)
Certificates of Analysis (1)
//dev-homepage/applications/digital-pcr/promotions/registration/search/products?query=QIAGEN%20Plasmid%20Mini%20Kit%20(100/search3/products?query=dna/search3/products/dev-search/products?query=dna/dev-search/products/product-categories/discovery-and-translational-research/genomic-services/product-categories/discovery-and-translational-research/genomic-services?disable-wfc=true&disable-dtm=true/products/products/diagnostics-and-clinical-research/sample-processing/allprep-rnaprotein-kit/products/diagnostics-and-clinical-research/sample-processing/allprep-rnaprotein-kit/products/discovery-and-translational-research/dna-rna-purification/dna-purification/genomic-dna/blood-and-cell-culture-dna-mini-kit/products/discovery-and-translational-research/pcr-qpcr/pcr-enzymes-and-kits/hifidelity-long-range-and-other-pcr/ucp-hifidelity-pcr-kit/products/discovery-and-translational-research/lab-essentials/buffers-reagents/maxtract-high-density/products/discovery-and-translational-research/lab-essentials/buffers-reagents/maxtract-high-density?catno=129056/products/discovery-and-translational-research/lab-essentials/buffers-reagents/maxtract-high-density?catno=129065/products/diagnostics-and-clinical-research/sample-processing/allprep-rnaprotein-kit/BAD_URL/product-categories/discovery-and-translational-research/genomic-services/BAD_URL/products?cmpid=1234&intcmp=456/products?cmpid=asdf&intcmp=qwertysds-searchpromotionsknowledge-and-support/resourcesapplications/enzymes/tools-and-calculatorsapplications/enzymes/tools-and-calculators/ligation-calculator